ARCHITECTi2000SR

ARCHITECT- i2000SR

ARCHITECT-i2000SR

The ARCHITECT i2000SR Immunoassay Analyzer meets your laboratory's high standards by delivering STAT results when you need them. The advanced technology and flexible protocols built into the ARCHITECT i2000SR enhance laboratory workflow and allow you to report results with confidence.

Specification

SPECIFICATIONS DESCRIPTIONS
MetdodsCHEMIFLEX
Maximum troughputUp to 200 tests/hour
Sample TypesSerum, Plasma, Whole Blood, Urine
Sample TubesHeight: 72-102 mm Diameter: 9.6 -16.1 mm
Sample CupYes (50 µl dead volume)
Sample Capacity65
Sample Barcode TypesCode 39, Codabar, Interleaved 2 of 5, Code 128
Sample Result Storage50,000
Sample Volume10-150 μL, Average: 50 µL
Automatic DilutionYes
Sample Probe Carryover<0.1 parts per million
Reagent Capacity25 refrigerated positions
Reagent Type100% liquid ready-to-use liquid reagents
Reagent Onboard StabilityMajority of assays are stable on board 30 days
Calibration FrequencyCalibrate witd new lot number,controls are specified otderwise
Sample Clot and Bubble DetectionYes
Reagent Pressure MonitoringYes
Sample Interference MeasurementNo
System Control Center1 SCC, witd color touchscreen , keyboard, and mouse
On-board Maintenance RecordsYes
Online Error Code HelpYes
Host InterfaceBidirectional, serial RS-232 interface, query option available
Remote DiagnosticsAbbottLink®
Dimensions (HxWxD)49” x 61” x 49” 121.9 x 154.9 x 124.5 cm
Weight1,081 lb 490.3 kg
Electrical RequirementsAC 180-264V±10%, 47-63 Hz
Water RequirementsPurified water to dilute buffer concentrate
Heat Output4280 BTU/hr, Running Mode
Get a Quote

Request A Quote